MAINTENANCE
8.1.
Maintenance Schedule
NOTE
The operations outlined in this chapter are to be
performed by qualified maintenance personnel only.
Table 8-1: Preventative Maintenance Schedule
Item
Maintenance Interval
Reference Section
TEST functions
6 - 12 month intervals
Zero/Span Calibration
Annually or after repairs
Zero/Span checks
Daily
Particulate Filter
Weekly as needed
Filter for PermaPure Drier
Replace every 24 months
Filter for Ozone Scrubber
Replace every 24 months
Drierite Canister
Replace when Drierite changes
from blue to red.
from blue to red.
Reaction cell window
Clean annually or as necessary
Ozone Flow
Check TEST function every
year
year
Sample Flow
Check TEST function every
year
year
Converter
Check efficiency every 6
months
months
Pneumatic Lines
Examine every 12 months,
clean if necessary
clean if necessary
Factory Calibration
Calibrate each year or after
repairs
repairs
Leak Check
Check every year or after other
maintenance.
maintenance.
Rx Cell O-rings
Replace every 12 months
O-rings
Replace every 12 months