Teledyne API Model 200AH NO
X
 Analyzer Instruction Manual,  06492, Rev. G2 
 
8-2 
8.2. 
  Replacing the Sample Particulate Filter 
The particulate filter should be inspected often for signs of plugging or contamination. It is also 
common for dirt particles to absorb NO
2
, thus causing the reading to be low. A discoloration of 
the filter membrane indicates that the filter is dirty and should be changed.  
To change the filter: 
1.  Locate the sample filter on the outside rear panel. 
2.  Unscrew the hold-down ring, remove the Teflon o-ring and then the filter disc. See 
Figure 8-1 for an exploded view of the filter assembly. 
3.  Replace the filter, being careful that the element is fully seated in the bottom of the holder. 
Replace the Teflon o-ring, then screw on the hold-down ring and hand tighten.