Teledyne API Model 200AH NO
X
 Analyzer Instruction Manual,  06492, Rev. G2 
 
5-21 
To set up Auto Range press SETUP-RNGE-MODE-AUTO, then press ENTR. To set the values 
press SETUP-RNGE-SET. The M200AH will prompt you for LO, then HI which is the lower 
and upper ranges of Auto Range. Key in the value desired, then press ENTR. 
Remote Range works the same as AutoRange except that the LO/HI range decision is made via 
remote status inputs on the rear panel, see Figure 2-2 for pinout. Calibration of Remote Range is 
also the same as for AutoRange see Section 7.8. To select Remote AutoRange press SETUP-
RNGE-MODE-REM, then press ENTR. 
5.3.4.3.   Independent Ranges (NO
X
 Measurements) 
Independent Ranges allows you to select different ranges for NO, NO
X
, and NO
2
To set up Independent Ranges press SETUP-RNGE-MODE-IND, then press ENTR. To set the 
values press SETUP-RNGE-SET. The M200AH will prompt you for the range of NO, NO
X
 and 
NO
2
 channels. Key in the desired range for each channel, then press ENTR after each value. 
5.3.4.4.   O
2
 Range 
To set the value for the range press SETUP-RNGE-SET-O2, and enter the full scale range 
desired from 5 percent to 100 percent. 
5.3.4.5.   Dilution Ratio 
The dilution feature allows the M200AH to be used with a stack dilution probe. With the 
Dilution feature you can select the range and display the concentration at the value and units of 
the un-diluted gas in the stack. The dilution probe dilutes the gas by a fixed ratio so the analyzer 
is actually measuring a much lower concentration than is actually present in the stack. 
The software scales the diluted sample gas concentration readings so that the outputs show the 
actual stack concentrations. Also, when calibrating the instrument or setting the ranges the 
values selected are scaled to reflect the actual stack concentrations. The scaled readings are sent 
to the display, analog outputs, and RS-232 port. 
To use the Dilution feature: 
1.  SELECT UNITS 
For stack measurement, select PPM units. To set units, press SETUP-RANGE-UNIT-PPM. Press 
ENTR after the unit selection is made, then EXIT to return to upper level menus. 
2.  SET DILUTION RATIO 
The dilution ratio of the probe is entered by SETUP-RANGE-DIL. Accepted values are 1 to 
1000. Press ENTR, and EXIT to return to upper level menus. A value of 1 disables the dilution 
feature.