Teledyne API Model 200AH NO
X
 Analyzer Instruction Manual,  06492, Rev. G2 
 
5-22 
3.  SELECT RANGE  
The range selection is the same with dilution as with normal monitoring. See Section 5.3.4 for 
information on range selection. You should note however, the value entered should be the actual 
concentration of the calibration gas entering the dilution probe. The units of this number is ppm. 
4.  CALIBRATION 
When the above selections have been made, the instrument now must be calibrated through the 
dilution probe. 
NOTE 
 
5.3.4.6.   Concentration Units 
The M200AH can display NO
X
 concentrations in ppm or mg/m
3
 and if equipped with the 
Oxygen Sensor option, displays O
2
 concentrations in %. To change the current NO
X
 units press 
SETUP-RANGE-UNIT. Select PPM or MGM, then press ENTR. Remember, you should now 
reset the expected span concentration values in the new units and re-calibrate the instrument 
using one of the methods in Section 7. The following equations give approximate conversion 
values. The exact values depend on the standard temperature and pressure used by your 
regulatory agency.  
NO ppm x 1.34 = NO mg/m
NO
2
 ppm x 2.05 = NO
2
 mg/m
 
NOTE 
Changing units affects all of the RS-232 values, all of the  
display values, and all of the calibration values. 
Example: 
If the current units are in ppm with the span value at 400 ppm, and  
the units are changed to mg/m
3
 the span value is NOT re-calculated  
to the equivalent value in mg/m
3
. Therefore the span value now  
becomes 400 mg/m
3
 instead of 400 ppm. Re-enter the expected  
span concentrations in mg/m3 units and recalibrate.