Teledyne 200AH 用户手册
Teledyne API Model 200AH NO
X
Analyzer Instruction Manual, 06492, Rev. G2
5-27
5.5.
4-20 mA Current Loop
The analog signal output is available on the rear panel as a voltage or 4-20 mA current output.
Optionally the current loop outputs can be isolated. See Section 6.3 for a description of isolated
current loop outputs.
Optionally the current loop outputs can be isolated. See Section 6.3 for a description of isolated
current loop outputs.
The current outputs come out on the same terminals that are used for voltage outputs see
Figure 2-2. The NO, NO
Figure 2-2. The NO, NO
x
, NO
2
and Test Channel outputs have current loop capabilities. See
The factory default is voltage outputs. To convert from voltage to current loop, move the
jumpers on the motherboard from the default B-C (rear pins) position to the A-B (forward pins)
position. See the 01110 Motherboard schematic in the Appendix.
jumpers on the motherboard from the default B-C (rear pins) position to the A-B (forward pins)
position. See the 01110 Motherboard schematic in the Appendix.
5.6.
Status Output
The status output is an option that reports Analyzer conditions via contact closures on the rear
panel. The closures are available on a 50 pin connector on the rear panel. The contacts are NPN
transistors which can pass 50 ma of direct current. The pin assignments are listed in Table 5-8.
panel. The closures are available on a 50 pin connector on the rear panel. The contacts are NPN
transistors which can pass 50 ma of direct current. The pin assignments are listed in Table 5-8.
Table 5-8: Status Output Pin Assignments
Output #
Pin #
Definition
Condition
1
1,2
ZERO CAL
CLOSED IN ZERO CAL
2
3,4
SPAN CAL
CLOSED IN HIGH SPAN CALIBRATION
3
5,6
FLOW ALARM
CLOSED IF FLOW WARNING
4
7,8
TEMP ALARM
CLOSED IF ANY TEMP WARNING
5
9,10
DIAG MODE
CLOSED IN DIAG MODE
6
11,12
POWER OK
CLOSED IF SYSTEM POWER OK
7
13,14
CONC ALARM
CLOSED IF CONCENTRATION
WARNING
WARNING
8
27,28
LOW SPAN CAL
CLOSED IN LOW SPAN CALIBRATION
9
SPARE
10
19,20
AUTORANGE - HI
CLOSED IF IN HIGH RANGE
11
21,22
SYSTEM OK
CLOSED IF NO FAULTS PRESENT
12
23,24
RX CELL PRESS
CLOSED IF ABS PRES > 15" HG
The Status Board schematic can be found in the Appendix Drawing 01087.