Teledyne API Model 200AH NO
X
 Analyzer Instruction Manual,  06492, Rev. G2 
 
5-27 
5.5. 
  4-20 mA Current Loop  
The analog signal output is available on the rear panel as a voltage or 4-20 mA current output. 
Optionally the current loop outputs can be isolated. See Section 6.3 for a description of isolated 
current loop outputs. 
The current outputs come out on the same terminals that are used for voltage outputs see  
Figure 2-2. The NO, NO
x
, NO
2
 and Test Channel outputs have current loop capabilities. See 
The factory default is voltage outputs. To convert from voltage to current loop, move the 
jumpers on the motherboard from the default B-C (rear pins) position to the A-B (forward pins) 
position. See the 01110 Motherboard schematic in the Appendix. 
5.6. 
  Status Output 
The status output is an option that reports Analyzer conditions via contact closures on the rear 
panel. The closures are available on a 50 pin connector on the rear panel. The contacts are NPN 
transistors which can pass 50 ma of direct current. The pin assignments are listed in Table 5-8. 
Table 5-8:  Status Output Pin Assignments 
Output # 
Pin # 
Definition 
Condition 
1,2 
ZERO CAL 
CLOSED IN ZERO CAL 
3,4 
SPAN CAL 
CLOSED IN HIGH SPAN CALIBRATION 
5,6 
FLOW ALARM 
CLOSED IF FLOW WARNING 
7,8 
TEMP ALARM 
CLOSED IF ANY TEMP WARNING 
9,10 
DIAG MODE 
CLOSED IN DIAG MODE 
11,12 
POWER OK 
CLOSED IF SYSTEM POWER OK 
13,14 
CONC ALARM 
CLOSED IF CONCENTRATION 
WARNING 
27,28 
LOW SPAN CAL 
CLOSED IN LOW SPAN CALIBRATION 
9  
SPARE 
 
10 
19,20 
AUTORANGE - HI 
CLOSED IF IN HIGH RANGE 
11 
21,22 
SYSTEM OK 
CLOSED IF NO FAULTS PRESENT 
12 
23,24 
RX CELL PRESS 
CLOSED IF ABS PRES > 15" HG 
 
The Status Board schematic can be found in the Appendix Drawing 01087.