Teledyne 200AH 用户手册
Teledyne API Model 200AH NO
X
Analyzer Instruction Manual, 06492, Rev. G2
5-35
5.7.3. TEST Commands and Messages
Table 5-15: RS-232 Test Messages
Name
Message
Description
NOXRANGE
1
NOX RNG=xxxxx ppm
Indep. range for NO
x
channel
NORANGE
(1)
NO RNG=xxxxx ppm
(3)
Indep. range for NO channel
NO2RANGE
(1)
NO2 RNG=xxxxx ppm
(3)
Indep. range for NO
2
channel
RANGE
(2)
RANGE=xxxxx
ppm
(3)
Analyzer
range
STABIL
STABIL=xxxx.xx ppm
Std. Deviation of last 10 min of conc values
FLOW
SAMPLE FL=xxx CC/M
Sample flow rate
OFLOW
OZONE FL=xxxx CC/M
Ozone flow rate
PMT
PMT=xxxxxx MV
PMT output
NORM PMT
NORM_PMT=xxxxxxx MV
PMT output adjusted for T/P compensation
AZERO
AZERO=xxxxxx MV
AutoZero offset value
HVPS
HVPS=xxxxx V
High voltage power supply
DCPS
DCPS=xxxxxx MV
DC power supply
RCTEMP
RCELL TEMP=xxx C
Reaction cell temperature
BOXTEMP
BOX TEMP=xxx C
Internal box temperature
PMTTEMP
PMT TEMP=xxx C
PMT temperature
BLOCKTEMP
BLOCK=xxx C
Sample flow control block temp
CONVTEMP
CONV TEMP=xxx C
Converter temperature
RCPRESS
RCEL=xxx.x IN-HG-A
Rx Cell Pressure
SMPPRESS
SAMP=xxx.x IN-HG-A
Sample Pressure
NOXSLOPE NOXSLOPE=xxxxx
NO
X
Slope Parameter
NOXOFFSET NOXOFFS=xxxxx
NO
X
Offset Parameter
NOSLOPE
NOSLOPE=xxxxxx
NO Slope Parameter
NOOFFSET
NOOFFS=xxxxxx
NO Offset Parameter
CLKTIME
TIME=HH:MM:SS
Time of day
(1)
Displayed when independent range is enabled.
(2)
Displayed when single or AutoRange is enabled.
(3)
Depends on which units are currently selected.