Teledyne 200AH 用户手册

下载
页码 196
Teledyne API Model 200AH NO
X
 Analyzer Instruction Manual,  06492, Rev. G2 
 
5-35 
5.7.3.   TEST Commands and Messages 
Table 5-15:  RS-232 Test Messages 
Name 
Message 
Description 
NOXRANGE
1
 
NOX RNG=xxxxx ppm
 
Indep. range for NO
x
 channel 
NORANGE
(1)
 
NO RNG=xxxxx ppm
(3)
 
Indep. range for NO channel 
NO2RANGE
(1)
 
NO2 RNG=xxxxx ppm
(3)
 
Indep. range for NO
2
 channel 
RANGE
(2)
 RANGE=xxxxx 
ppm
(3)
 Analyzer 
range 
STABIL 
STABIL=xxxx.xx ppm 
Std. Deviation of last 10 min of conc values 
FLOW 
SAMPLE FL=xxx CC/M 
Sample flow rate 
OFLOW 
OZONE FL=xxxx CC/M 
Ozone flow rate 
PMT 
PMT=xxxxxx MV 
PMT output 
NORM PMT 
NORM_PMT=xxxxxxx MV 
PMT output adjusted for T/P compensation 
AZERO 
AZERO=xxxxxx MV 
AutoZero offset value 
HVPS 
HVPS=xxxxx V 
High voltage power supply 
DCPS 
DCPS=xxxxxx MV 
DC power supply 
RCTEMP 
RCELL TEMP=xxx C 
Reaction cell temperature 
BOXTEMP 
BOX TEMP=xxx C 
Internal box temperature 
PMTTEMP 
PMT TEMP=xxx C 
PMT temperature 
BLOCKTEMP 
BLOCK=xxx C 
Sample flow control block temp 
CONVTEMP 
CONV TEMP=xxx C 
Converter temperature 
RCPRESS 
RCEL=xxx.x IN-HG-A 
Rx Cell Pressure 
SMPPRESS 
SAMP=xxx.x IN-HG-A 
Sample Pressure 
NOXSLOPE NOXSLOPE=xxxxx 
 
NO
X
 Slope Parameter 
NOXOFFSET NOXOFFS=xxxxx 
 
NO
X
 Offset Parameter 
NOSLOPE 
NOSLOPE=xxxxxx  
NO Slope Parameter 
NOOFFSET 
NOOFFS=xxxxxx 
NO Offset Parameter 
CLKTIME 
TIME=HH:MM:SS 
Time of day 
 
(1)
Displayed when independent range is enabled. 
(2)
Displayed when single or AutoRange is enabled. 
(3)
Depends on which units are currently selected.