The T command lists TEST messages. Examples of the T command are: 
T LIST  
Lists test messages currently enabled with T SET 
T LIST ALL 
Lists all test messages 
T CONVTEMP 
Prints the temperature of the converter 
T LIST NOX 
Prints NO
x
 concentration message 
T NOX 
Prints concentration message 
 
5.7.4.   WARNING Commands and Messages 
Table 5-16:  RS-232 Warning Messages 
Name 
Message 
Description 
WSYSRES 
SYSTEM RESET 
Analyzer was reset/powered on 
WRAMINIT 
RAM INITIALIZED 
RAM was erased 
WSMPFLOW 
SAMPLE FLOW WARN 
Sample flow out of spec. 
WOZONEFLOW 
OZONE FLOW WARNING 
Ozone flow out of spec. 
WRCELLP 
RCELL PRESS WARNING 
Rx cell pressure out of spec. 
WBOXTEMP 
BOX TEMP WARNING 
Box temp. out of spec. 
WRCELLT 
RCELL TEMP WARNING 
Reaction cell temp. out of spec. 
WCONVTMP 
CONV TEMP WARNING 
Converter temp. out of spec. 
WOBLKTMP 
ORIFICE BLK TEMP WARN 
Orifice block temp out of spec. 
WDCPS 
DC POWER SUPPLY WARN 
DC Power voltage out of spec. 
WDYNZERO 
CANNOT DYN ZERO 
Dynamic zero cal. out of spec. 
WDYNSPAN 
CANNOT DYN SPAN 
Dynamic span cal. out of spec. 
WBLKTEMP 
SAMPLE FLOW ORIFICE 
BLOCK TEMP WARNING  
Temperature out of spec. 
WOZONEGEN 
OZONE GEN OFF 
Ozone generator is off 
WAUTOZERO 
AZERO WRN XXX.X MV 
AutoZero reading out of spec. 
WADINS 
A/D NOT INSTALLED 
A/D board not installed or broken