Teledyne API Model 200AH NO
X
 Analyzer Instruction Manual,  06492, Rev. G2 
 
7-3 
7.1. 
  Manual NO
X
 Zero/Span Check or Cal With Zero/Span 
Gas in the Sample Port 
The calibration of the instrument can be checked or adjusted using gas coming in the sample 
port. This method is often used when the calibration gas is supplied from the stack sample probe 
or external tank and valve system. This is the calibration method to use if the instrument is 
purchased without the Zero/Span Valve. 
Table 7-2:  Manual Zero Calibration Procedure - Zero Gas Thru Sample Port 
Step Number 
Action 
Comment 
1. Check 
“Before 
Calibration” 
2. 
Press CAL 
The M200AH enters the calibrate mode from sample mode. 
The zero gas must come in through the sample port. 
3. 
Select LO or HI 
Range 
If AutoRange is enabled you will be prompted for which 
range to calibrate. Select the LO or Hi range, then press 
ENTR. 
4. 
Wait 10 min 
Wait for reading to stabilize at zero value. 
5. 
Press ZERO 
If you change your mind after pressing ZERO, you can still 
press EXIT here without zeroing the instrument. 
6. 
Press ENTR 
Pressing ENTR actually changes the calculation equations. 
7. 
Press EXIT 
M200AH returns to sampling. Immediately after calibration, 
data is not added to the DAS averages.