  CALIBRATION AND ZERO/SPAN CHECKS 
There are several ways to check and adjust the calibration of the M200AH. These different 
methods are summarized in the following Table 7-1. In addition, all of the methods described in 
this section can be initiated and controlled via the RS-232 port, see Section 5.7 for details. 
We strongly recommend that NO
X
 SPAN CALIBRATION be done with NO span gas. Since NO 
is unaffected by the converter, both channels can be accurately calibrated. It is also important 
that there be some oxygen in the span gas. Inaccurate calibration of the NO
X
 channel can result 
from calibration gasses containing no oxygen. All calibration gasses must have a dew point of 
< 20
C. 
NOTE 
The quality of calibration can be no better than the accuracy and  
 
NOTE 
If there are any problems completing any of the following procedures,  
 
Before Calibration 
1.  Use the SETUP-RNGE menu to set the range the instrument will be operated in. 
2.  If you are using any of the following features, they must be set up before calibration: 
A.  Autoranging or Remote Ranging - See Section 7.8 
C.  NO only or NO
X
Normal calibration assumes that these features are already set up.