Teledyne API Model 200AH NO
X
 Analyzer Instruction Manual,  06492, Rev. G2 
 
7-6 
7.2. 
  Manual O
2
 Zero/Span Check or Cal With Zero/Span 
Gas in the Sample Port 
The calibration of the O
2
 Sensor can be checked or adjusted using gas coming in the sample port. 
This method is often used when the calibration gas is supplied from the stack sample probe or 
external tank and valve system.  
Table 7-5 details the zero calibration procedure with zero gas coming in through the sample port. 
Table 7-5:  Manual O
2
 Zero Calibration Procedure - Zero Gas Thru Sample Port 
Step Number 
Action 
Comment 
1. Check 
“Before 
Calibration” 
2. Press 
 
CAL-O2-ENTR 
The M200AH enters the calibrate mode for O
2
. The zero gas 
must come in through the sample port. 
4. 
Wait 10 min 
Wait for reading to stabilize at zero value. 
5. 
Press ZERO 
If you change your mind after pressing ZERO, you can still 
press EXIT here without zeroing the instrument. 
6. 
Press ENTR 
Pressing ENTR actually changes the calculation equations. 
7. 
Press EXIT 
M200AH returns to sampling. Immediately after calibration, 
data is not added to the DAS averages. 
 
Table 7-6 details the span calibration procedure with span gas coming in through the sample 
port.