Teledyne API Model 200AH NO
X
 Analyzer Instruction Manual,  06492, Rev. G2 
 
7-7 
Table 7-6:  Manual O
2
 Span Calibration Procedure - Span Gas Thru Sample Port 
Step Number 
Action 
Comment 
1. Press 
CAL-O2-CONC 
This key sequence causes the M200AH to prompt for the 
expected O
2
 concentration. 
 
Enter the O
2
 span concentration value by pressing the key 
under each digit until the expected value is set.  
2. 
Press ENTR 
Pressing ENTR stores the O
2
 span value and returns the 
prompt to the CAL menu. 
3. 
Wait 10 min 
Wait for reading to stabilize at span value. 
4. 
Press SPAN 
If you change your mind after pressing SPAN, you can still 
press EXIT here without spanning the instrument. 
5. 
Press ENTR 
Pressing ENTR actually changes the calculation equations 
and causes the instrument to read the expected O
2
 span 
concentration. 
6. 
Press EXIT 
M200AH returns to sampling. Immediately after calibration, 
data is not added to the DAS averages. 
 
7.3. 
  Manual Zero/Span Check with Zero/Span Valves 
Option 
The Zero/Span valve option can be operated from the front panel keyboard as described below. 
The Zero/Span valves allow zero and span gas come in through ports on the rear panel of the 
instrument. The Zero/Span valves can be set up to operate automatically on a periodic schedule. 
See Section 6.2 to setup automatic zero/span feature.