2-5 ELECTRICAL CONNECTIONS 
NOTE 
Electrical connnections must be made in 
compliance with National Electrical Code 
(ANSI/NFPA 70) and/or any applicable na-
tional or electical codes. 
Two electrical connections are required on the 
Analyzer Module: POWER and NETWORK 
(See Figure 2-4 above).  On the Analyzer Mod-
ule, two NETWORK connectors are available, 
either of which is appropriate for: 1) intercon-
nection with the Backplane of the Platform or 2) 
"daisy-chaining" with other NGA2000 compo-
nents (A star connection is acceptable for LON 
lengths under about 10 meters.) 
Connect a source of 24 V 5A DC power to the 
power inlet.  Make sure that the ground connec-
tion is made, and that this is separate from the 
power return lead.  Failure to ensure a good 
ground may result in random noise and distur-
bance in the analyzer readings.