Teledyne T100 Benutzerhandbuch

Seite von 422
 
 
29 
2. 
SPECIFICATIONS, APPROVALS & COMPLIANCE 
This section presents specifications for the T100 analyzer and the O
2
 and CO
2
 sensor 
options, Agency approvals, EPA equivalency designation, and CE mark compliance. 
2.1. SPECIFICATIONS AND APPROVALS 
Table 2-1  T100 Basic Unit Specifications 
Parameter Description 
Ranges  
(Physical Analog Output) 
Min: 0-50 ppb Full Scale 
Max: 0-20,000 ppb Full Scale (selectable, dual ranges and auto ranging supported) 
Measurement Units 
ppb, ppm, µg/m
3
, mg/m
3
 (selectable) 
Zero Noise
1
 
0.2 ppb (RMS) 
Span Noise
1
 
0.5% of reading, above 50 ppb 
Lower Detectable Limit
2
 0.4 
ppb 
Zero Drift 
0.5 ppb/24 hours 
Span Drift  
0.5% of full scale/24 hours 
Lag Time
1
 20 
seconds 
Rise/Fall Time
1
  
<100 sec to 95% 
Linearity 
1% of full scale 
Precision
1
 
0.5% of reading above 50 ppb 
Sample Flow Rate 
650 cm
3
/min. ±10% 
Power Requirements 
100V-120V, 60Hz (165W); 220V-240V, 50 Hz (140W) 
Analog Output Ranges 
10 V, 5 V, 1 V, 0.1 V (selectable) 
Recorder Offset 
± 10 % 
Standard I/O 
1 Ethernet: 10/100Base-T 
2 RS-232 (300 – 115,200 baud)  
2 USB device ports 
8 opto-isolated digital outputs 
6 opto-isolated digital inputs 
4 analog outputs 
Optional I/O 
1 USB com port 
1 RS485 
8 analog inputs (0-10V, 12-bit) 
4 digital alarm outputs 
Multidrop RS232 
3 4-20mA current outputs 
06807C DCN6650