Teledyne 460L Benutzerhandbuch

Seite von 92
TELEDYNE INSTRUMENTS
 
 
Serial Communications 
460L Instruction Manual 
30 
05228 Rev B 
 
PRINTED DOCUMENTS ARE UNCONTROLLED 
DCN 5164 
7.3.2. ALMSTAT 
SYNTAX 
<address>ALSTAT<CR> 
DESCRIPTION 
Returns the current status of both of the concentration alarms.   
0= Inactive 
1= Active 
DATA PARAMETERS SENT 
None 
RESPONSE 
<address>:<HI alarm state>,<HI-HI alarm state>#<checksum><CR> 
EXAMPLE 
Command:   
1ALMSTAT<CR> 
Response: 
1:0,0#247<CR> 
- Both concentration alarms are INACTIVE 
1:1,0#248<CR> 
- The HI alarm is ACTIVE and the HI-HI alarm is INACTIVE 
1:1,1#249<CR> 
- Both concentration alarms are ACTIVE