Laser LS8000-3 Benutzerhandbuch

Seite von 221
LaserSpeed 8000-3 Instruction Handbook
 
RS-232/RS-422 Communication 
 
 
Part No. 93463 / Drawing No. 0921-01561 
Page 163 of 221 
Revision A (Sep 2007) 
 
*AUTOADDR 
UDP Power-On Data Destination IP Address
*AUTOADDR?<CR>
 
 
Read 
  Response 
Format: 
*AUTOADDR=<N><CR>
 
 
 
Set 
*AUTOADDR=<XXX.XXX.XXX.XXX><CR> 
 
Where  <XXX.XXX.XXX.XXX> is an IP Address, such as "192.168.10.246".  This 
command configures the destination IP address for UDP output data.  If the gauge is not 
configured to output UDP data on start-up, then this setting is not used. 
 
*AUTOPORT 
UDP Power-On Data Destination Port
*AUTOPORT?<CR>
 
 
Read 
  Response 
Format: 
*AUTOPORT=<N><CR>
 
 
 
Set 
*AUTOPORT=<N><CR> 
 
Where  <N> is an IP port number, such as 1234.  This command configures the destination 
Port for UDP output data.  If the gauge is not configured to output UDP data on start-up, then 
this setting is not used.  This port should be set to a value higher than 1024 so that it doesn't 
interfere with standard network interfaces.