ANCB Installation  
Description 
The ANCB is installed into slot 1 of the Advance Maxum’s or NAU’s 
System Controller (SYSCON). To install the ANCB it is not necessary to 
disconnect interface cables or to remove the SYSCON from the unit. 
 
Slot 1 
Voltage dangerous to life exists. Before performing the removal and 
installation procedures, it is important that primary AC power to the 
unit be turned off from the main circuit breaker. Observe all plant 
safety requirements before performing any repair or maintenance 
on any unit  
WARNING 
 
 
NAU Installations 
For rack mount NAU units the SYSCON connections are visible at the 
rear of the NAU. If you have a wall mounted unit you must first remove 
the front or side safety box access covers to see the SYSCON 
connections. 
 
Instructions 
Step Procedure 
1.   
Locate backplane Slot 1, and remove the screw that holds 
the metal strip in place. Keep the screw. 
      
 
SLOT 1 
Screw 
Metal 
Plate 
 
2000592-001 
Installation 