Fluke 124 Benutzerhandbuch
![Fluke](https://files.manualsbrain.com/attachments/9d57315a76f85bd9af23da46bf7ecaf7f248b87d/common/fit/150/50/7671c6748e9e70e4c5afb082a772388950496a1161e6cfdedf4b942e284b/brand_logo.png)
Specifications
Safety
4
73
Multimeter disturbance:
•
VDC, VAC, and VAC+DC with STL120 and short ground lead.
•
OHM, CONT, DIODE, and CAP with STL120, and black test lead to COM.
Table 3
Disturbance less than 1% of full scale
E= 3 V/m
E= 10 V/m
Frequency range 10 kHz to 27 MHz
VDC, VAC, VAC+DC
OHM, CONT, DIODE
CAP
OHM, CONT, DIODE
CAP
500 mV to 1250V
500
Ω to 30 MΩ
50 nF to 500
µF
500 mV to 1250V
500
Ω to 30 MΩ
50 nF to 500
µF
Frequency range 27 MHz to 1 GHz
VDC, VAC, VAC+DC
OHM, CONT, DIODE
CAP
OHM, CONT, DIODE
CAP
500 mV to 1250V
500
Ω to 30 MΩ
50 nF to 500
µF
500 mV to 1250V
500
Ω to 30 MΩ
50 nF to 500
µF
Test tool ranges not specified in table 3 may have a disturbance of more than 10% of full scale.