Overview of running a study 
          Calibrating 
053-644, Rev E  
Page   34 
 
eg4 & eg3 
 
34 
3M Quest Technologies recommends to calibrate your dosimeter before you run your study(ies) 
and after your study(ies) to ensure highly accurate measurement results.  The steps below 
address how to calibrate for either a pre or post calibration.   
 
 
Calibrating the dosimeter 
 
1.  Remove the windscreen from the dosimeter by unscrewing it.  (See Figure 6-1 for an 
illustration, page 50.) 
 
2.  Ensure the Edge is turned on and in the stop mode.  (See “Turning on” page 8, and “Turning 
Off” page 9). 
 
3.  Slide the Edge into the mouth of the calibrator. 
 
NOTE: ensure the adapter ring is inserted into the mouth of the calibrator before you slide the 
Edge into the cavity.  
 
4.  Turn on the calibrator and ensure parameters are set to 114 dB and 1000 Hz. (See calibration 
manual for operation details.) 
 
 
 
 
 
 
 
 
 
 
 
     
Figure 4-4:  Calibrating 
 
5.  “CAL” will flash on the display when detected. 
 
6.  Press the on/off enter         button to trigger the calibration display. 
 
7.  To match the values on the calibration display, press repeatedly the up or down arrow in 
order to select the exact output on the calibrator (i.e., 114 dB). 
 
8.  When you have selected the appropriate dB level, press the on/off enter           button.  
 
 
A shaded bar will appear loading your calibration information.  Then a summary calibration 
screen will appear indicating the db level with date and time displayed.  
  
9.  To return to the measurement view screen, press the on/off enter            . 
   
10. To end the calibration steps, turn off the calibrator.  Detach the dosimeter from the calibrator.   
“CAL” flashes