Overview of running a study 
          Setting contrast on dosimeter 
053-644, Rev E  
Page   36 
 
eg4 & eg3 
 
36 
 
Contrast 
 
1.  Ensure the meter is Off
 
2.  Turn the meter on by pressing the on/off/enter           button.   
 
3.  When the splash screen appears, press and hold the run/stop          and on/off enter        
buttons. (See figure 4-5 for splash screen).      
 
4.  The Setup menu appears on the Edge.  Repeatedly press the arrow key until the Contrast 
symbol is selected.   
 
NOTE:  Setup menu is used to reset the dosimeter, change the contrast, and return to the 
main view. 
 
 
 
 
 
 
 
 
 
 
 
                                  Figure 4-7:  Changing the contrast from the keypad 
 
5.  Next press the on/off enter          button. 
 
6.  A number will appear on your display (i.e., 50).  Press the up or down arrow to increase or 
decrease the contrast of the Edge. 
 
7.  Press on/off enter          button to accept the new contrast value. 
 
8.  To return to the measurement view screen, press the down arrow and press the            
on/off/enter           button. 
Checkmark icon 
Is the selection to 
exit the setup menu 
Contrast icon 
 
Arrows and enter key 
Indicates the valid keys 
are selected