The Interface Screen
Figure 3.4 – Interface Screen
Key Features of the Interface Screen - This screen allows you to set the serial
communication baud rate for the port used. It also allows for the use of Ethernet, which
includes UDP and TCP/IP port setup.
Display of Communication Events on the Main Menu – Check this box to view
communication events on the Main Menu screen.
Hide Patient Name Received From Host for Privacy – If this box is checked, the patient’s
name will not be shown during processing.
Run 30 Second Loopback Test – Check this box to run a loopback test on the serial port.
This test verifies that the port can send and receive data. You must insert a loopback plug in
the port to perform this test. Press the test button. In the box below the button, you will see a
string of letters and numbers if the port is working. You will not see any characters in the box
if the port is not functioning.
communication baud rate for the port used. It also allows for the use of Ethernet, which
includes UDP and TCP/IP port setup.
Display of Communication Events on the Main Menu – Check this box to view
communication events on the Main Menu screen.
Hide Patient Name Received From Host for Privacy – If this box is checked, the patient’s
name will not be shown during processing.
Run 30 Second Loopback Test – Check this box to run a loopback test on the serial port.
This test verifies that the port can send and receive data. You must insert a loopback plug in
the port to perform this test. Press the test button. In the box below the button, you will see a
string of letters and numbers if the port is working. You will not see any characters in the box
if the port is not functioning.
Communications Log – This button is a diagnostic tool that may be used by Technical
Support.