Clarity User Guide,
 
March 1, 2011 - Rev. 1.06 
                    
           Page 24
 
The Interface Screen 
 
 
Figure 3.4 – Interface Screen 
 
Key Features of the Interface Screen - This screen allows you to set the serial 
communication baud rate for the port used.  It also allows for the use of Ethernet, which 
includes UDP and TCP/IP port setup.  
 
Display of Communication Events on the Main Menu – Check this box to view 
communication events on the Main Menu screen. 
 
Hide Patient Name Received From Host for Privacy – If this box is checked, the patient’s 
name will not be shown during processing.  
 
Run 30 Second Loopback Test – Check this box to run a loopback test on the serial port.  
This test verifies that the port can send and receive data. You must insert a loopback plug in 
the port to perform this test.  Press the test button.  In the box below the button, you will see a 
string of letters and numbers if the port is working.  You will not see any characters in the box 
if the port is not functioning. 
 
 
Communications Log – This button is a diagnostic tool that may be used by Technical 
Support.