How to Measure Blood Pressure 
(in MANU. Mode) 
Wrap the arm cuff over the 
patient’s arm.
1.
Push the ON/OFF (power) 
Button to turn on the power.
3.
ON/OFF
AVG./1st/2nd/3rd
DEFLATION
START
STOP
HIDE
P-SET
100
AUTO
140
180
220
260
280
MODE
AVG.
MANU.
CHECK
SINGLE
Set the P-SET (pressure set-
ting) Volume to “AUTO” or the 
target in ation value.
When setting the P-SET to 
“AUTO”, turn the Volume coun-
terclockwise as far as it goes un-
til you can hear the click sound.
When the systolic pressure is 
expected to exceed 220 mmHg, 
proper in ation may not be pos-
sible in “AUTO” setting.
Set the target in ation value to 
30 to 40 mmHg higher than the 
expected systolic pressure.
5.
ON/OFF
AVG./1st/2nd/3rd
DEFLATION
START
STOP
HIDE
P-SET
100
AUTO
140
180
220
260
280
MODE
AVG.
MANU.
CHECK
SINGLE
Push the ON/OFF (power) 
Button to turn off the power.
7.
ON/OFF
AVG./1st/2nd/3rd
DEFLATION
START
STOP
HIDE
P-SET
100
AUTO
140
180
220
260
280
MODE
AVG.
MANU.
CHECK
SINGLE
Place the stethoscope on the 
patient’s arm.
2.
ON/OFF
AVG./1s
t /2nd/3r
d
DEFLATION
START
STOP
HIDE
P-SET
100
AUTO
140
180
220
260
280
MODE
AVG.
MANU.
CHECK
SINGLE
Set the MODE Selector to 
“MANU.”.
4.
ON/OFF
AVG./1st/2nd/3rd
DEFLATION
START
STOP
HIDE
P-SET
100
AUTO
140
180
220
260
280
MODE
AVG.
MANU.
CHECK
SINGLE
Push the START Button to 
start the measurement.
In the “AUTO” setting, the unit 
stops in ation at the pres-
sure considered to be 30 to 
40 mmHg above the expected 
systolic pressure, then start 
de ation.
The unit can be re-in ated 
only while the START Button is 
pushed.
The pressure display is syn-
chronized with the pulse rate 
and renewed.
Each time the DEFLATION 
(de ation control) Button is 
pushed, the unit de ates by 
5~10 mmHg.
When you press the STOP But-
ton, the unit de ates fast and 
ends the operation.
6.
ON/OFF
AVG./1st/2nd/3rd
DEFLATION
START
STOP
HIDE
P-SET
100
AUTO
140
180
220
260
280
MODE
AVG.
MANU.
CHECK
SINGLE