How to Measure Blood Pressure 
(in CHECK Mode)
Display of pressure can be con rmed by the CHECK Mode (yearly procedure).
What you need to prepare
(1) Well-adjusted reference pressure monitor (including rubber ball), 
(2) T-shaped tube, 
(3) Two rubber tubes, and (4) something in the cylindrical shape to wrap the arm cuff such as the tea container (sturdy one 
that will not break or deform with pressure)
* Measurement error may occur even with the mercury blood pressure monitor because of insuf cient amount of mercury 
in the mercury, dirty glass tube, or clogged glass tube cap.
How to use the blood pressure monitor 
Connect the mercury blood 
pressure monitor, rubber 
ball, and this unit with the 
T-shaped tube as shown in 
the Figure.
Push the ON/OFF (power) 
Button to turn on the 
power.
Set the MODE Selector to 
“CHECK”.
Tightly wrap the arm cuff 
over a sturdy cylindrical 
object.
Close the air release valve 
of rubber ball and in ate 
the monitor to a certain 
value (pressure to be 
checked).
Compare the value displayed on this monitor and that on the mercury.
Open the air release valve of rubber ball to de ate.
CHECK result
(1) In the measuring rule, the test tolerance of blood pressure monitor (accuracy of pressure when shipped from the 
factory) should not exceed 3 mmHg.
(2) If the CHECK result is +/- 3 mmHg or above, contact the dealer where you purchased this unit or the nearest OMRON 
dealer.