National Instruments NI PXI-5422 Benutzerhandbuch

Seite von 63
NI PXI-5422 Calibration Procedure
16
ni.com
Verifying the Oscillator Frequency Accuracy
This test verifies the frequency accuracy of the oscillator on the 
NI PXI-5422. The verification involves generating a 10 MHz sine wave 
with the NI PXI-5422 and measuring the sine wave frequency with one of 
the instruments from Table 2.
To verify the frequency accuracy of the oscillator on the NI PXI-5422, 
complete the following steps:
1.
Connect the NI PXI-5422 CH 0 front panel connector to the instrument 
measuring the frequency accuracy with a male BNC to female SMB 
cable.
2.
Call 
niFgen_init
 (niFgen Initialize VI) using the following 
parameters:
resourceName: The name of the device that you want to verify. 
You can find this name under Devices and Interfaces in MAX.
IDQuery
VI_TRUE
resetDevice
VI_TRUE
vi: A pointer to a ViSession. The variable passed by reference 
through this parameter receives the value that identifies the 
session created by this function. This value acts as the session 
handle and is passed as the first parameter to all subsequent 
NI-FGEN functions.
3.
Call 
niFgen_SetAttributeViReal64
 to set the sample rate 
(niFgen property node: Arbitrary Waveform Output»Sample Rate
using the following parameters:
vi: The session handle returned from 
niFgen_init
channelName
"0"
attributeID
NIFGEN_ATTR_ARB_SAMPLE_RATE
value
200000000
4.
Call 
niFgen_SetAttributeViReal64
 to set the gain (niFgen 
property node: Arbitrary Waveform Output»Arbitrary Waveform 
Gain
) using the following parameters:
vi: The session handle returned from 
niFgen_init
channelName
"0"
attributeID
NIFGEN_ATTR_ARB_GAIN
value
1
 
Note 
You can adjust this value based on which measuring device you use.