National Instruments NI PXI-5422 Benutzerhandbuch

Seite von 63
NI PXI-5422 Calibration Procedure
18
ni.com
A frequency error of 45 Hz for a 10 MHz signal corresponds to an 
error of 4.5 ppm. This limit accounts for the initial accuracy and the 
frequency deviation caused by temperature and aging. Refer to Table 3 
for frequency ranges.
11. Call 
niFgen_AbortGeneration
 (niFgen Abort Generation VI) to 
abort the waveform generation using the following parameter:
vi: The session handle returned from 
niFgen_init
12. Call 
niFgen_close
 (niFgen Close VI) to close the instrument 
driver session, to destroy the instrument driver session and all of 
its properties, and to release any memory resources NI-FGEN uses. 
Use the following parameter:
vi: The session handle returned from 
niFgen_init
Verifying the DC Gain and Offset Accuracy
This test verifies the DC gain and offset accuracy of the NI PXI-5422 into 
a high-impedance load by generating a number of DC voltages and offsets, 
measuring the voltage with a DMM, and comparing the NI PXI-5422 to the 
error limits.
The DC gain and offset accuracy verification procedure has three 
sub-procedures that verify the following:
Main analog path gain
Main analog path offset
Direct path gain
Verifying the Main Analog Path Gain
To verify the gain of the NI PXI-5422 main analog path, complete the 
following steps:
1.
Connect the NI PXI-5422 CH 0 front panel connector to the DMM for 
measuring DC gain and offset accuracy.
2.
Call 
niFgen_init
 (niFgen Initialize VI) using the following 
parameters:
resourceName: The name of the device that you want to verify. 
You can find this name under Devices and Interfaces in MAX.
Table 3.  Frequency Ranges
Calibration Test Limit
Published Specifications ±25 ppm
Low
High
Low
High
9,999,955 Hz
10,000,045 Hz
9,999,750 Hz
10,000,250 Hz