CON 400/410 
- 14 - 
Figure 9 - TDS Conversion Factor 
From measurement mode 
 
1. Press 
Setup key to enter Set Up mode. 
 
2. Press 
the 
S and T keys to scroll through subgroups 
until you view parameter P4.0. 
 
See Figure 9. 
 
3. Press 
the 
ENTER key again. The upper display shows 
a value and the lower display shows “tdS”. 
 
4.  Calculate the TDS factor of your solution. See 
Addendum 2 on page 46 for information on how to 
calculate the TDS factor. 
 
5. Press 
the 
S and T keys to select your calculated TDS 
conversion factor. 
 
6. Press 
the 
ENTER key to confirm selection and to 
return to the subgroup menu. Press the CAL/MEAS 
key to return to measurement mode. 
 
 
 
 
 
 
 
 
ppt
ppm
SETUP
p 4.4
ppt
ppm
SETUP
0.50
MR
MI
ENT
ER
RA
NG
E
SETUP
cof
p 4.0
MR
MI