eutech-instruments waterproof cyberscan con 400410 & tds 400 conductivitytds meter Manuale Utente
Instruction Manual
CON 400/410
- 14 -
Figure 9 - TDS Conversion Factor
From measurement mode
1. Press
Setup key to enter Set Up mode.
2. Press
the
S and T keys to scroll through subgroups
until you view parameter P4.0.
See Figure 9.
3. Press
the
ENTER key again. The upper display shows
a value and the lower display shows “tdS”.
4. Calculate the TDS factor of your solution. See
Addendum 2 on page 46 for information on how to
calculate the TDS factor.
5. Press
the
S and T keys to select your calculated TDS
conversion factor.
6. Press
the
ENTER key to confirm selection and to
return to the subgroup menu. Press the CAL/MEAS
key to return to measurement mode.
ppt
ppm
ppm
SETUP
p 4.4
ppt
ppm
ppm
SETUP
0.50
MR
MI
ENT
ER
RA
NG
E
SETUP
cof
p 4.0
MR
MI