- 17 - 
5.2.2 
Setting a manual temperature compensation value  
To use manual temperature compensation, you need to 
manually enter the temperature value of your process into 
the meter. You can select any temperature between 0 and 
100 
°C (32 to 212 °F). Default value is 25 °C.  
 
To select a manual temperature compensation value 
1. 
Switch the meter on. Press the MODE key to select 
measurement mode.  
2. 
If necessary, select ATC off as described in section 
5.2.1 on page 16.  The ATC indicator will not appear 
on the display.  
3. Press 
the 
CAL/MEAS key to enter conductivity or TDS 
calibration mode. The CAL indicator will appear above 
the primary display.  
4. 
While in conductivity or TDS calibration mode, press 
the MODE key to enter temperature calibration mode. 
The primary display shows the current temperature 
setting and the secondary display shows the default 
value 25 
°C (77 °F) or its last set value.  
5. 
Check the temperature of your sample using an 
accurate thermometer.  
6. Press 
the 
MI/
S or MR/T keys to offset the 
temperature to the measured value from step 5.  
7. Press 
ENTER to confirm the selected temperature and 
to return to the conductivity or TDS measurement 
mode.  
 
The meter will now compensate conductivity or TDS readings for manually set temperature.  
 
NOTES: 
To exit this program without confirming the manual temperature compensation value, DO NOT 
press ENTER in step 7.  Press CAL/MEAS instead.  
 
Figure 9 - Manual Temperature 
Compensation