5.4 
Using Manual Ranging Function 
Your meter automatically selects the range in which your readings appear. The manual 
ranging function lets you select the specific range (and corresponding resolution) that you 
want to work in:  
CON 400 and CON 410 meters: 
CON 410 meter: 
1. 
0.00 --- 19.99 
µS 
1. 
0.00 --- 9.99 ppm 
2. 
0.0 --- 199.9 
µS 
2. 
0.0 --- 99.9 ppm 
3. 
0 --- 1999 
µS 
3. 
0 --- 999 ppm 
4. 
0.00 --- 19.99 mS 
4. 
0.00 --- 9.99 ppt 
5. 
0.0 --- 199.9 mS 
5. 
0 --- 100 ppt  
(up to 199.9 ppt depending on TDS factor 
setting) 
1. 
To select the desired measuring range, press the 
RANGE key while in Measurement mode. The first 
range will appear on the display and the “MEAS” 
indicator blinks.  
2. Press 
the 
RANGE key again (if needed) until desired 
range is selected.  
3. 
To re-select the Auto-ranging function, repeatedly 
press the RANGE key until the “MEAS” indicator 
appears without blinking.  
 
NOTES 
If the value of the solution you are measuring is higher than 
the range selected “Or” will appear on the primary display. 
Press RANGE until the correct range is selected.  
The meter resets to the Auto-ranging function once it is 
turned off. You will have reset the manual ranging function 
each time you turn the meter off.  
 
Figure 11 - Manual ranging 
Figure 12 - Out-of-range