Instruction Manual
CON 400/410
- 27 -
7.2
P 1.0: Memory Clear (CLr)
Use this parameter to clear all stored memory values when you need to store a new series of
values. This lets you avoid confusing the old values with the new ones.
NO is the default setting.
NOTE: Selecting YES will wipe out all memory.
From measurement mode:
1. Press
the
SETUP key to enter Set Up mode.
2. Press
the
MI/
S or MR/T keys to scroll through
subgroups until you view the parameter P1.0.
3. Press
the
ENTER key to enter parameter P1.0.
4. Press
the
MI/
S or MR/T keys to toggle between
NO and YES. NO retains current memory; YES
clears all memory. See Figure 23.
5. Press
the
ENTER key to confirm selection and
return to the subgroup menu. Press CAL/MEAS key
to return to measurement mode.
Figure 23 - P1.0: Memory Clear