eutech-instruments waterproof cyberscan con 400410 & tds 400 conductivitytds meter Benutzerhandbuch

Seite von 56
Instruction Manual 
 
CON 400/410 
- 29 - 
7.4 
P3.0: Viewing probe data 
Program 3 has five “view only” options that let you check 
the probe’s parameters for diagnostic purposes.  These 
options show you the effective cell constant for each range. 
The cell constant is adjusted according to your calibration.  
 
From measurement mode 
1. Press 
the 
SETUP key to enter Set Up mode.  
2. Press 
the 
MI/
S or MR/T keys to scroll through 
subgroups until you view parameter P3.0.   
3. Press 
the 
ENTER key repeatedly to view the effective 
cell constant for each range.  
4. 
When you have scrolled through all probe data, you 
will automatically return to the subgroup menu. Press 
the CAL/MEAS key to return to measurement mode.  
See Figure 25.  
 
 
 
 
NOTES:  
Cell constants may degrade with time and usage. You 
can use this feature to alert you to the need for a new 
probe prior to total failure.  
 
 
Figure 25 - View probe data for 
each measurement range