CON 400/410 
- 33 - 
 
7.5.4 
P4.4  Setting the TDS factor 
For CON 410 meter only 
The concentration of salts dissolved in solution increases the conductivity of that solution. This 
relationship varies from salt to salt and is roughly linear over a given range for a given salt.  
The TDS conversion factor is the number used by the meter to convert from conductivity to 
TDS.  
 
To determine the conductivity to TDS conversion 
factor for your solution: 
Addendum 2 on page 46 explains the calculation in 
determining the conductivity to TDS conversion factor. 
You can set the TDS conversion factor between 0.40 and 
1.00; meter default is 0.50.  
 
From measurement mode 
1. Press 
SETUP key to enter Set Up mode.  
2. Press 
the 
MI/
S or MR/T keys to scroll through 
subgroups until you view parameter P4.0.  
3. Press 
the 
ENTER key five times to select parameter 
4.4. The upper display shows “tdS” and the lower 
display shows “P4.4”.  
4. Press 
the 
ENTER key again. The upper display 
shows a value and the lower display shows “tdS”.  
5. 
Calculate the TDS factor of your solution. See 
Addendum 2 on page 46 for information on how to 
calculate the TDS factor.  
6. Press 
the 
MI/
S or MR/T keys to select your 
calculated TDS conversion factor.  
7. Press 
the 
ENTER key to confirm selection and to 
return to the subgroup menu. Press the CAL/MEAS 
key to return to measurement mode.  
 
 
Figure 29 - Change of TDS factor