Instruction Manual 
 
CON 400/410 
- 35 - 
7.6.2 
P5.2  Adjusting the normalization temperature 
Your meter will normalize its conductivity measurements to a standard temperature that you 
can select. You can adjust the normalization temperature from 15 to 30 
°C (59 to 86 °F). Meter 
default is 25 
°C (77 °F).  
 
From measurement mode 
1. Press 
SETUP key to enter Set Up mode. 
2. Press 
the 
MI/
S or MR/T keys to scroll through 
subgroups until you view parameter P5.0.  
3. Press 
the 
ENTER key three times to select 
parameter 5.2. The display shows “t.nr” on the 
upper display.  
4. Press 
the 
ENTER key again. The upper display 
shows the normalisation temperature and the lower 
display shows “t.nr”. 
5. Press 
the 
MI/
S or MR/T keys to select the 
normalisation temperature.  
6. Press 
the 
ENTER key to confirm selection and to 
return to the subgroup menu. Press CAL/MEAS key 
to return to measurement mode.  
 
 
NOTE: 
Most calibration standards are referenced to 25°C. If a 
normalization temperature other than 25°C is entered, the 
calibration standard value at this temperature must be 
known. 
 