9. Depress the syringe plunger to completely empty the insulin 
into the Pod. The Pod will beep, indicating that the System is 
ready to proceed to the next step.
10. Remove the needle from the insulin fill port. The port is self-
sealing; insulin will not leak after the needle is removed.
11. Place the protective cap back on the needle and remove the 
needle from the syringe
.
12. After you fill the Pod, it will beep twice. After hearing the 
beeps, PDM and pod should be adjacent and touching, 
either in or out of tray
 to ensure proper communication 
during priming (Figure 5-7, on the next page). Press Next.
13. The Pod will only beep if you have filled it with at least 85 units 
of insulin. If you have filled the Pod with more than 85 units 
and still do not hear the 2 beeps, call Customer Care.
14. The System performs a series of safety checks and automati-
cally primes the Pod (Figure 5-8). Once complete, the PDM 
beeps, letting you know that the priming and safety checks 
were successful.
To ensure proper fill, do not insert the fill syringe 
at an angle into the fill port.
Do not use any other type of needle or filling 
device besides the syringe provided with each 
Pod.
NEVER use a Pod if you hear a crackling noise or 
feel resistance when you depress the plunger. 
These conditions can result in interrupted insulin 
delivery of insulin.
Do not insert the fill syringe into the fill port more 
than once.
NEVER inject air into the fill port. Doing so may 
result in unintended or interrupted insulin deliv-
ery.
The fill syringe is intended for single use only and 
should be used only with the OmniPod System.
As a safety feature, the communication distance be-
tween the Pod and PDM is reduced during activation. 
Once a Pod is primed and communicates with the 
PDM, the full communication range is restored and 
the Pod can receive commands only from that PDM.