Insulet Corporation 017 Benutzerhandbuch

Seite von 190
46
5
4. Determine the amount of insulin you need to insert into the 
Pod. For example, if you will use this Pod for 48 hours, you 
need enough insulin to last you 48 hours. Your healthcare 
provider will help you determine the correct amount.
5. Draw air into the syringe up to the amount of insulin you 
want.
6. Insert the needle into the insulin vial and inject the air. This 
makes it easier to withdraw insulin from the vial.
7. Turn the vial and syringe upside down. Withdraw insulin from 
the vial into the syringe, expelling any and all air or air bub-
bles. Fill at least to the MIN (minimum) fill line (Figure 5-5)
Refer to the Care Giver Guide as an additional reference.
8. Remove the needle from the vial and insert it straight down 
into the insulin fill port on the underside of the Pod 
(Figure 5-6).
Use care after removing the needle cap and 
exposing the fill needle.
Only use room temperature insulin when filling 
the Pod.
The Pod requires a minimum of 85 units of insulin to 
begin operation.
The Pod can deliver up to 200 units of insulin.
Make sure there is no air in the syringe before 
attempting to fill a Pod with insulin.
Be sure to only insert the fill syringe into the fill 
port. Attempting to inject insulin into any other 
location on the Pod may result in damage to the 
Pod or loss of insulin.
Figure 5-5
Figure 5-6
Fill
port