6
a. To edit an existing target BG time segment, choose the 
segment, then press Edit. Use the Up/Down Controller 
buttons to choose a new start time, end time, target BG 
value, and correct-above value (correction threshold). Press 
Next after each entry.
b. To add a segment, choose [add new], then press New
Use the Up/Down Controller buttons to choose a start 
time, end time, target BG value, and correct-above value 
(correction threshold). Press Next after each entry.
c. When you have completed all the time segments you 
want, press Done, then Save.
2. Min BG for calcs (smallest blood glucose value allowed for 
calculating boluses)—Use the Up/Down Controller buttons 
to enter a new minimum BG, then press Next.
3. Insulin-to-carb (IC) ratio—You can edit existing time seg-
ments and add segments, up to a total of 8.
a. To edit an existing IC ratio time segment, choose the seg-
ment, then press Edit (Figure 6-7 on next page). Use the 
Up/Down Controller buttons to choose a new start time, 
end time, and IC ratio. Press Next after each entry.
b. To add a segment, choose [add new], then press New
Use the Up/Down Controller buttons to choose a start 
time, end time, and IC ratio. Press Next after each entry.
c. When you have completed all time segments, press 
Done, then Save.
 
4. Correction factor—You can edit existing time segments and 
add segments, up to a total of 8.
a. To edit an existing correction factor time segment, choose 
the segment, then press Edit. Use the Up/Down Control-
ler
 buttons to choose a new start time, end time, and cor-
rection factor. Press Next after each entry. 
b. To add a segment, choose [add new], then press New
Use the Up/Down Controller buttons to choose a start 
time, end time, and correction factor. Press Next after 
each entry.
c. When you have completed all time segments, press 
Done, then Save.
Figure 6-7