Baumer Electric AG OIS-W-WR300303 Benutzerhandbuch

Seite von 74
Baumer Ident AG 
 
12 / 13 
 
Author: Clemens Zehnder   
 
File: Maintenance-SW.doc Version 3.0 e     
 
 
4.1.3  Page 3: Operation Mode Settings 
 
Picture 4-3: operation mode settings 
Name Description 
Range 
Default 
Remarks 
Mmain 
main operation mode 
0..1 
0=event driven,1=triggered 
Mtxuntrig 
transmitter state when not triggered;  
for triggered and event driven mode 
0..1 1 
0=off,1=on 
Mbiased biased 
mode 
0..1 
0  0=off,1=on 
Mrandom 
random mode; ramp timing distributions 
0..3 
0=off, 1=uniform distributed, 2,3=not yet released 
MRTm 
max. measurements per block 
0..255 
25 
the min measurement block corresponds to 
Parameter „Reader Settings: Nequ“. 
Also used for triggered mode when random = OFF. 
MRNmess  number of measurement blocks 
1..255 5 
 
MRTdmin 
min. random delay 
1..65535 
in ramp durations
1
 
MRTdlen 
range of random delay 
1..65535 
50 
in ramp durations 
Mtsleep 
sleep time 
1..65535 
50 
in ramp durations 
Table 4-1: operation mode settings 
                                                      
1 Ramp duration =  20 ms