GENERAL CARE
WARNING: Always clean and/or disinfect reusable sensors and accessories between patients to reduce the
risk of spreading infectious agents between persons.
risk of spreading infectious agents between persons.
Maintenance
The following table shows the recommended maintenance procedures for the S2, patient accessories, and Li-Ion Battery
Charger. The S2 should be serviced once a year by a Mortara authorized service technician; however, it is good practice
to periodically ensure the telemetry device is in proper working order. This can be performed by a clinician or biomed
at the hospital or healthcare delivery organization familiar with the S2 telemetry device, ECG signal acquisition, as well
as general maintenance/calibration of biomedical equipment.
Charger. The S2 should be serviced once a year by a Mortara authorized service technician; however, it is good practice
to periodically ensure the telemetry device is in proper working order. This can be performed by a clinician or biomed
at the hospital or healthcare delivery organization familiar with the S2 telemetry device, ECG signal acquisition, as well
as general maintenance/calibration of biomedical equipment.
To accomplish these steps in their entirety and verify the correct operation of the system, appropriate patient simulators
or other equipment may be required. Refer to the service manual for further details.
or other equipment may be required. Refer to the service manual for further details.
Functionality
Procedure
Mechanical Integrity
Check for cracks, abrasive edges and other signs of damage.
ECG Telemetry
Connect ECG leads to Patient Simulator.
Start a new monitoring session.
Verify that waveforms for all leads are properly shown on the LCD.
Li-Ion Battery Charger
Check for cracks, abrasive edges and other signs of damage.
Check that all connectors and AC cord length is unbroken and smooth along its
length.
length.
Verify proper LED indicators during battery charging.
ECG Cables
Approved Cleaning Agents
Enzymatic detergent such as ENZOL (US) or CIDEZYME (outside the US).
Distilled water.
Disinfectant solution (such as CIDEX OPA, or a 10% solution of household
bleach (5.25% sodium hypochlorite) in distilled water).
bleach (5.25% sodium hypochlorite) in distilled water).
Soft, lint-free cloths and/or soft-bristled brushes.
Protective gloves and eyewear.
Procedure
1. Remove the battery from the telemetry device.
2. Put on gloves and protective eyewear.
3. Prepare the detergent according to the manufacturer's instructions, and also
2. Put on gloves and protective eyewear.
3. Prepare the detergent according to the manufacturer's instructions, and also
the disinfectant solution, in separate containers.
4. Apply detergent to product using a soft, lint-free cloth. If material is dried on,
allow to sit for 1 minute. Do not immerse cable ends or lead wires in liquid as
it can cause corrosion.
it can cause corrosion.
5. Wipe smooth surfaces with the cloth.
6. Use a soft-bristle brush on visibly soiled areas and irregular surfaces.
7. Remove detergent from product using cloth dampened in distilled water.
8. Repeat as necessary.
9. Apply disinfectant solution on affected area using a soft cloth. Allow product to
6. Use a soft-bristle brush on visibly soiled areas and irregular surfaces.
7. Remove detergent from product using cloth dampened in distilled water.
8. Repeat as necessary.
9. Apply disinfectant solution on affected area using a soft cloth. Allow product to
sit for 5 minutes.
10. Wipe excess solution and clean product again with cloth dampened in distilled
water.
11. Allow 2 hours for drying.
WARNING: Only Mortara authorized service providers should perform servicing of the S2 to ensure that the
correct maintenance and calibration procedures are followed and that the S2 is returned to proper operation.
Refer to Section 1 for a list of technical support and service providers.