Inventek Systems 341 User Manual

Page of 63
19 of 63 
 
 
 
GENERAL CARE
 
 
WARNING: Always clean and/or disinfect reusable sensors and accessories between patients to reduce the 
risk of spreading infectious agents between persons. 
 
Maintenance
 
The following table shows the recommended maintenance procedures for the S2, patient accessories, and Li-Ion Battery 
Charger. The S2 should be serviced once a year by a Mortara authorized service technician; however, it is good practice 
to periodically ensure the telemetry device is in proper working order. This can be performed by a clinician or biomed 
at the hospital or healthcare delivery organization familiar with the S2 telemetry device, ECG signal acquisition, as well 
as general maintenance/calibration of biomedical equipment. 
 
To accomplish these steps in their entirety and verify the correct operation of the system, appropriate patient simulators 
or other equipment may be required.   Refer to the service manual for further details. 
 
Functionality
 
Procedure
 
Mechanical Integrity 
Check for cracks, abrasive edges and other signs of damage. 
ECG Telemetry 
 
Connect ECG leads to Patient Simulator. 
 
Start a new monitoring session. 
 
Verify that waveforms for all leads are properly shown on the LCD. 
Li-Ion Battery Charger  
 
Check for cracks, abrasive edges and other signs of damage. 
 
Check that all connectors and AC cord length is unbroken and smooth along its 
length. 
 
Verify proper LED indicators during battery charging. 
ECG Cables 
Approved Cleaning Agents 
 
Enzymatic detergent such as ENZOL (US) or CIDEZYME (outside the US). 
 
Distilled water. 
 
Disinfectant solution (such as CIDEX OPA, or a 10% solution of household 
bleach (5.25% sodium hypochlorite) in distilled water). 
 
Soft, lint-free cloths and/or soft-bristled brushes. 
 
Protective gloves and eyewear. 
 
Procedure 
1.  Remove the battery from the telemetry device. 
2.  Put on gloves and protective eyewear. 
3.  Prepare the detergent according to the manufacturer's instructions, and also 
the disinfectant solution, in separate containers. 
4.  Apply detergent to product using a soft, lint-free cloth. If material is dried on, 
allow to sit for 1 minute. Do not immerse cable ends or lead wires in liquid as 
it can cause corrosion. 
5.  Wipe smooth surfaces with the cloth. 
6.  Use a soft-bristle brush on visibly soiled areas and irregular surfaces. 
7.  Remove detergent from product using cloth dampened in distilled water. 
8.  Repeat as necessary. 
9.  Apply disinfectant solution on affected area using a soft cloth. Allow product to 
sit for 5 minutes. 
10.  Wipe excess solution and clean product again with cloth dampened in distilled 
water. 
11.  Allow 2 hours for drying. 
 
WARNING: Only Mortara authorized service providers should perform servicing of the S2 to ensure that the 
correct maintenance and calibration procedures are followed and that the S2 is returned to proper operation. 
Refer to Section 1 for a list of technical support and service providers.