    Fusion Wireless Recorder 
 
15 
Record ECG:  ECG to be marked will be described by the physician.   
1.  Press the RECORD/ENTER button until an audible tone is heard, 
then release.   
2. 
Hold as still as possible during recording, which should only last a 
few seconds, but continue breathing.  
 
 
Automatic Recording:   
If an event is detected, the Recorder will silently record and transmit the 
event to the monitoring center for further review.   
 
Note about TEXT Messaging:   
The Fusion Wireless Monitoring System Software can provide text 
messaging back the Fusion Recorder.  Messages up to 3 lines with 32 
characters per line can be displayed on the LCD of the Fusion Recorder 
to allow communications back to the patient.  A TEXT message received 
by a recorder will initiate an audible alert of three beeps in rapid 
succession.  The alert will repeat every 10 minutes until the patient 
presses one the arrow keys or is silenced by the monitoring center.  The 
message shall be displayed continuously until cleared by the monitoring 
center.  The patient is not able to clear the message unless the batteries 
are removed.
 
 
To Send: 
In most cases, events are automatically downloaded to the receiving 
center via digital cellular link.  If an appropriate cellular signal is not 
present for the transmission to start, the recorder will automatically 
connect to the Home Link Bluetooth modem. There isn’t any patient 
interaction required for this transmission to occur.   
 
To Download via USB: 
Once the recording is finished, remove the patient cable from the 
recorder.  Connect the recorder to a PC running at the service center.  
Software on the PC will download the ECG as needed.  There is no 
patient interaction with USB and USB cables cannot be connected while 
a patient cable is connected. 
 
 
 
 
      
    Fusion Wireless Recorder 
 
16 
 
Troubleshooting (page 1 of 2) 
Symptom 
Recommended Solution 
No display 
Ensure batteries are inserted with correct polarity. 
Ensure RECORD button has been pressed. 
Will not record 
Ensure Patient Cable is inserted completely. 
Recorder stops recording