6) You are now back in the Set-Point Menu. Select “End” and push “EN-
TER” to return to normal Dyna-Link operation or select the other Set-
Point for additional modification as needed. (CLEAR/CANCEL will also
return the unit to normal operation).
UNIT(S)
Units can be changed from the front panel with the following procedure. Once
the units menu is enabled, units are changed by scrolling through the choices
with the UP (PEAK) key or DOWN (ZERO) key and pushing ENTER
(POWER) when the desired unit is displayed. Available units are: lb (pounds),
kg (kilograms), t (short tons), daN (dekaNewtons), and kg X1000 (equivalent to
Metric tons).
To change the units
1) Turn off the power. While holding down the “PEAK” key, turn the power
back on. After 1 second release the “PEAK” key.
2) The display reads “Unit” (if not push the UP or DOWN key to select the
Unit mode).
3) Push ENTER (POWER). You are now in the units menu and the current
units will be flashing.
4) Push UP/DOWN until the desired units are displayed, then push ENTER
(POWER).
5) Now you are back in the front panel Setup Menu. To go back to normal
operation, push CLEAR/CANCEL (ZERO and PEAK pushed simulta-
neously or “*” key on Remote units) or scroll to the “End” message and
push “ENTER”.
SETUP/CAL
Function:
Allows front panel entry of seldom set parameters and initiates Calibration.
SETUP/CAL is invoked by pushing the key through the access hole in the back
plate of the Dyna-Link (Remote units only: the SETUP/CAL key is located on
the front board inside the remote hand-held housing). For most of the setup
functions the Dyna-Link’s keys will work differently until SETUP/CAL is
exited. The POWER key will operate as an ENTER (POWER) key, the ZERO
key will operate as the DOWN Selection Key, the PEAK key will operate as
the UP Selection Key and the SETUP/CAL key will be used as a CLEAR/
CANCEL key. CLEAR/CANCEL can also be initiated via the front keys by
pushing ZERO and PEAK at the same time.