MSI-7200 Dyna-Link   
   User Guide     Page 7
ZERO SWITCH
Function:
Sets the Zero reading of the Dyna-Link.
Rules for Use:
1) The Dyna-Link must be stable within 2 display counts. If the reading is
not stable, the zero button will not function.
2) The Dyna-Link will accept a zero setting up to full capacity. Zero
settings will subtract from the overall capacity of the Dyna-Link. For
example, if you zero out 1000 lbs on a 5000 lb Dyna-Link, the overall
capacity will be reduced to 4000 lbs (plus the allowed over-range amount
- usually 9 divisions).
Action:
Push ZERO to zero the reading.
Final:
The digits display “0” (or 0.0, 0.00, or 0.000 depending on the resolution).
PEAK
Function:
Toggles between the Peak Tension and the Gross Tension modes.
Action:
1) Push PEAK.
2) Display annunciator “PEAK” lights indicating Peak Tension mode. The
“NEW PEAK” LED flashes while new peak values are being recorded.
3) The last stored peak reading is displayed.
Final:
The LCD digits update only when a higher peak reading is obtained When a
new higher reading is obtained, the “NEW PEAK” LED will flash four times.
To Clear the Peak
Action:
While in the Peak Hold Tension mode, push ZERO.
Final:
The old Peak value is cleared, the digits display the current gross tension as a
new peak. The “NEW PEAK” LED flashes to indicate a new peak being stored.
To Turn Off the Peak Mode
Action:
Push PEAK. The “PEAK” annunciator will go out.
Final:
The previous peak reading is retained even in the normal reading mode and