2.4.1  General introduction on how to connect a Pod to an instrument 
 
Follow the preceding steps to connect a Pod to an instrument. Disassemble the Pods by reversing the 
steps.   
 
 
1.  When inserting the Pod, keep the interface 
facing upwards.  
2.  Look for the tab when inserting the Pod. Begin 
by inserting the tab in the matching slot.  
3.  Clip the other side of the Pod on the 
instrument until the spring holder engages. 
 
Note: The system will automatically monitor 
improper physical connections between the Pods 
and the instruments 
 
 
2.4.2  Instrument Assembly for Tibia Cut First 
 
 
 
1.  Clip 1 x reference Pod (identifiable with a cross ‘X’ 
on the back cover) on the Tibial Alignment Guide 
[20-8011-013-00] 
 
2.  Clip 2x Pods (one cut guide and one 
reference) on Tibial Left or Right Adjustment 
mechanism (left: 20-8011-017-00, or right: 
20-8011-017-00), depending on operating 
leg. 
For Tibia Right, use the Pod identified by a 
square  
For Tibia Left, use the Pod identified by a