Orthosoft Inc. R20S21A Benutzerhandbuch
SmartKnee Surgical Technique
13 of 58
2.4.1 General introduction on how to connect a Pod to an instrument
Follow the preceding steps to connect a Pod to an instrument. Disassemble the Pods by reversing the
steps.
steps.
1. When inserting the Pod, keep the interface
facing upwards.
2. Look for the tab when inserting the Pod. Begin
by inserting the tab in the matching slot.
3. Clip the other side of the Pod on the
instrument until the spring holder engages.
Note: The system will automatically monitor
improper physical connections between the Pods
and the instruments
2.4.2 Instrument Assembly for Tibia Cut First
1. Clip 1 x reference Pod (identifiable with a cross ‘X’
on the back cover) on the Tibial Alignment Guide
[20-8011-013-00]
[20-8011-013-00]
2. Clip 2x Pods (one cut guide and one
reference) on Tibial Left or Right Adjustment
mechanism (left: 20-8011-017-00, or right:
20-8011-017-00), depending on operating
leg.
mechanism (left: 20-8011-017-00, or right:
20-8011-017-00), depending on operating
leg.
For Tibia Right, use the Pod identified by a
square
For Tibia Left, use the Pod identified by a