Orthosoft Inc. R20S21A Benutzerhandbuch
SmartKnee Surgical Technique
15 of 58
1. Clip 1x reference Pod (identified with a cross
‘X’ on the back cover) on the Femoral
Reference [20-8011-026-00]
Reference [20-8011-026-00]
2. Clip 2x Pods (one cute guide and one reference)
on Tibial Left or Right Adjustment mechanism
(left: 20-8011-017-00, or right: 20-8011-017-00),
depending on the operated leg
(left: 20-8011-017-00, or right: 20-8011-017-00),
depending on the operated leg
For Tibia Right, use the Pod identified by a
square
For Tibia Left, use the Pod identified by a circle
Note: The Tibial Adjustment Mechanism (left: 20-
8011-017-00 or right: 20-8011-017-00) is used even
during the Distal Femoral Cut because the system
uses this instrument to detect the patient side
8011-017-00 or right: 20-8011-017-00) is used even
during the Distal Femoral Cut because the system
uses this instrument to detect the patient side
9. Insert the Femoral Saw Slot [20-8011-042-00]
into the Femoral Distal Cut Guide [20-8011-
027-00]
027-00]
10. Tighten it with a 3.5 mm hex screw driver
(4812-45 or equivalent)
11. Clip 1x cut guide Pod on Femoral Distal Cut Guide
[20-8011-027-00]
Before executing this step, only one POD should
remain available with either the circle or square
marking
marking