Mindray DS USA Inc. WT0707-00000 Benutzerhandbuch
System Alarms Tab
System Setup Functions
9 - 22
0070-10-0690-01
Panorama™ Operating Instructions
System Alarm Limits
Select the System Alarm Limits sidebar button to display the System Alarm Limits
tab. For additional information regarding system alarm limits, refer to ‘‘System Alarms (Alarm
Limits) Tab’’ on page 9-23.
tab. For additional information regarding system alarm limits, refer to ‘‘System Alarms (Alarm
Limits) Tab’’ on page 9-23.
NOTE:
The System Alarm Limits sidebar button is not available at
the ViewStation or the WorkStation.
the ViewStation or the WorkStation.
System Alarm Responses
Select the System Alarm Responses sidebar button to display the System Alarm
Responses tab. For information regarding system alarm responses, refer to ‘‘Accessing the
System Alarms (System Alarm Responses) Tab’’ on page 9-26.
Responses tab. For information regarding system alarm responses, refer to ‘‘Accessing the
System Alarms (System Alarm Responses) Tab’’ on page 9-26.
NOTE:
The System Alarm Responses sidebar button is not available
at the ViewStation or the WorkStation.
at the ViewStation or the WorkStation.
Events Filtering
This sidebar button was previously described on page 9-18.
Normal Screen
Select the Normal Screen sidebar button to return the display to the full screen mode, and
close all tabs, message boxes, and dialog boxes.
close all tabs, message boxes, and dialog boxes.
9.6.2.6
Troubleshooting
This section lists some of the potential messages and issues that may occur while using the
System Alarm Options tab.
System Alarm Options tab.
MESSAGE/ISSUE *
REASON
SOLUTION
The System Setup menu
button is disabled.
button is disabled.
The System Setup function is
already open on a Panorama
display and cannot be opened
on multiple displays.
already open on a Panorama
display and cannot be opened
on multiple displays.
Close the System Setup
function on the Panorama
display on which it is opened.
function on the Panorama
display on which it is opened.
The Paging Delay button is
not displayed.
not displayed.
The central station where the
patient tile resides does not
have the Paging option
enabled.
patient tile resides does not
have the Paging option
enabled.
If the Paging feature was
purchased, enable it.
purchased, enable it.
The System Alarm Limits
and System Alarm
Responses tab is not
available at the ViewStation or
WorkStation.
and System Alarm
Responses tab is not
available at the ViewStation or
WorkStation.
System alarm limits and
responses must be changed at
the central station.
responses must be changed at
the central station.
Change the alarm limits and
responses from the central
station.
responses from the central
station.
Suspend All Timeout button
is disabled.
is disabled.
The Suspend Alarms option is
set to OFF or Non-Lethal.
set to OFF or Non-Lethal.
Change the Suspend Alarms
option to All.
option to All.
*
Messages are shown in all bold text.