System Alarms Tab
System Setup Functions
9 - 22
0070-10-0690-01
Panorama™ Operating Instructions
System Alarm Limits
Select the System Alarm Limits sidebar button to display the System Alarm Limits 
tab. For additional information regarding system alarm limits, refer to ‘‘System Alarms (Alarm 
Limits) Tab’’ on page 9-23.
NOTE:
The System Alarm Limits sidebar button is not available at 
the ViewStation or the WorkStation.
System Alarm Responses 
Select the System Alarm Responses sidebar button to display the System Alarm 
Responses 
tab. For information regarding system alarm responses, refer to ‘‘Accessing the 
System Alarms (System Alarm Responses) Tab’’ on page 9-26.
NOTE:
The System Alarm Responses sidebar button is not available 
at the ViewStation or the WorkStation.
Events Filtering
This sidebar button was previously described on page 9-18.
Normal Screen 
Select the Normal Screen sidebar button to return the display to the full screen mode, and 
close all tabs, message boxes, and dialog boxes.
9.6.2.6
Troubleshooting
This section lists some of the potential messages and issues that may occur while using the 
System Alarm Options tab.
MESSAGE/ISSUE * 
REASON
SOLUTION
The System Setup menu 
button is disabled.
The System Setup function is 
already open on a Panorama 
display and cannot be opened 
on multiple displays.
Close the System Setup 
function on the Panorama 
display on which it is opened.
The Paging Delay button is 
not displayed.
The central station where the 
patient tile resides does not 
have the Paging option 
enabled. 
If the Paging feature was 
purchased, enable it. 
The System Alarm Limits 
and System Alarm 
Responses
 tab is not 
available at the ViewStation or 
WorkStation.
System alarm limits and 
responses must be changed at 
the central station.
Change the alarm limits and 
responses from the central 
station.
Suspend All Timeout button 
is disabled.
The Suspend Alarms option is 
set to OFF or Non-Lethal.
Change the Suspend Alarms 
option to All.
*
Messages are shown in all bold text.