Operator’s Manual  ZM-520PA/521PA/530PA/531PA
 89
Specifications
ZM-520PA/530PA
Measuring Parameters
Measuring waveforms: 
ECG, respiration in impedance method, pulse
Measuring numeric data: 
Heart rate, respiration rate, SpO
2
, pulse rate
Transmitting Data
Waveform data: 
ECG, respiration, pulse wave
Numeric data: 
SpO
2
, pulse rate
Status information: 
Battery replacement, battery level*, alarm suspended, 
pause monitoring*, patient confirmed*, ECG lead, pacing 
detection, electrode detachment, electrode impedance*, 
ECG off*, respiration method (impedance)*, SpO
2
 status, 
respiration status*, channel ID, time constant (3.2 s), type 
of transmitter, transmitter code number*, transmitter serial 
number*
* The items marked with * are transmitted only when the protocol is “51”.
Display
Display size: 
2.2 inch TFT color LCD
Viewing area: 
44.16 (H) × 33.12 (V) mm
Resolution: 
320 (H) × 240 (V) dots
Displayed Data
Numeric and waveform screen: 
ECG (one waveform from lead I, II, III, Va or Vb), heart 
rate, pulse rate, respiration rate, SpO
2
, message, battery 
level, QRS/pulse sync mark, pulse bar graph, ECG lead
Waveform review screen: 
ECG or pulse wave of past 10 minutes
Numeric review screen: 
Heart rate or pulse rate, respiration rate and SpO
2
 at 1 
minute interval for past 10 minutes
Check electrodes screen: 
ECG for checking electrode attachment
ECG
ECG measurement
Channels: 4
Input dynamic range: 
±10 mV or more
Electrode offset potential tolerance:  ±500 mV or more
Input impedance: 
5 MΩ or more