Nihon Kohden Corporation ZM-520PA Benutzerhandbuch

Seite von 139
92 
Operator’s Manual  ZM-520PA/521PA/530PA/531PA
Response time:
 
Selectable from “Slow”, “Normal” and “Fast”.
 
The following graphs show the response time example when 
SpO
2
 changes 0.6%/s.
Pulse Rate = 70 bpm
60
65
70
75
80
85
90
95
100
0
10
20
30
40
50
60
70
80
90
100
110
120
130
140
Time (seconds)
Sp
O
2
 (%)
SpO
2
 Reference
SpO
2
 Fast
SpO
2
 Normal
SpO
2
 Slow
60
65
70
75
80
85
90
95
100
0
10
20
30
40
50
60
70
80
90
100
110
120
130
140
SpO
2
 Reference
SpO
2
 Fast
SpO
2
 Normal
Pulse Rate = 140 bpm
Time (seconds)
Sp
O
2
 (%)
SpO
2
 Slow
The following graph shows the response time example when pulse rate changes 10 bpm/s.