Operator’s Manual  ZM-520PA/521PA/530PA/531PA
 25
Turning Off the Transmitter
To turn off the power, remove batteries. When the power is turned off, the saved waveform and 
numeric data are deleted.
Changing the Transmitter Channel
The channel of the transmitter can be changed with an optional QI-901PK channel writer.
WARNING
The following actions must be taken to properly receive the transmitter signal of the 
correct patient on the receiving instrument. Otherwise, there may be signal loss or 
signals may mix causing a serious accident, such as monitoring a different patient.
•  Assign a channel administrator in the hospital and only he or she should manage 
channel assignment.
•  The channel administrator must manage the channels in the facility so that there is 
no signal interference.
•  When the transmitter channel is changed, the channel administrator must check 
that the channel on the receiving monitor is also changed and the signal is 
properly received.
•  The channel administrator must replace the channel number label on the 
transmitter with the new one after changing the channel.
NOTE
•  The software version of the QI-901PK channel writer must be 02-01 or later to change 
the channel on the transmitter.
•  The channel writer must be used outside the patient environment.
The channel is displayed in the upper left corner of the screen.
Channel