Nihon Kohden Corporation ZM-520PA Benutzerhandbuch

Seite von 139
Operator’s Manual  ZM-520PA/521PA/530PA/531PA
 27
Setting item
Description
Settings
SpO
2
 RESPONSE
Select the SpO
2
 response mode.
FAST: 
Select this for special applications 
that require a fast response. FAST is 
suitable for detecting short apnea.
NORMAL:  For normal monitoring.
SLOW: 
Select this when you need to suppress a 
rapid change in SpO
2
.
FAST, NORMAL, 
SLOW
Displaying the PARAMETER SETUP Screen
1.  Turn off the transmitter by removing one 
battery.
2.  While pressing the Function key, turn on the 
transmitter (install the battery). The MENU 
screen appears.
3. Press 
the 
▼ key to move the cursor to 
“PARAMETER SETUP”.
4.  Press the Function key to enter PARAMETER 
SETUP. The current settings are highlighted.
5. Change 
settings:
•  To move the cursor and select the setting 
item, press the 
 or 
▼ key then press the 
Function key.
•  To select and register the setting, press the 
 or 
▼ key then press the Function key.
•  To cancel changing the setting of the 
selected item, press the Screen key.
6.  When changing settings on the PARAMETER 
SETUP screen is complete, press the Screen 
key to return to the MENU screen. 
7. Press 
the 
 or 
▼ key to move the cursor to “EXIT”.
8.  Press the Function key. The numeric and waveform screen appears.
Cursor
MENU screen
Cursor
MENU screen
Setting item
PARAMETER SETUP screen - page 1
Setting
Cursor Setting item
PARAMETER SETUP screen - page 1
Setting
Cursor